• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number ROTAFLOW DRIVE,BLUE
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 11/22/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).
 
Event Description
(b)(4).It was reported, that during priming the error message "head error" was displayed when reaching 1000rpm.No clinical consequences were required as no patient was involved.
 
Event Description
Internal reference: (b)(4).Autonumber: (b)(6).
 
Manufacturer Narrative
The defective rotaflow drive was sent to the supplier em-tec (germany).According to the service report# rma#(b)(4) dated on (b)(6)2019 following was found: the failure could be reproduced.The defective ila 213 was replaced as well as the closing assy was replaced, oil on mast holder refiled.Burnin, calibration and final check system test were performed according to the service protocol.The unit passed all tests.The most probable root cause could be determined as mishandling and/or aging.Thus the failure could be confirmed.In the course of the investigation of nc-17-06-011 all complaints were analyzed individually to look for patterns and causes.After evaluation of the complaints, the following defects are the most common: hot plug, sig error followed by head error, error message due to shaking / error message due to sensitivity, connection problems and hardware-error.The increase in complaints with the error "head error" is due to a user / application error.The actions have been addressed in the past to prevent application errors.Furthermore, the instructions for use of the rotaflow system, see rotaflow system user manual, mcv-ga-10000703-de-11, contain detailed descriptions to prevent an "error head".In addition, the instruction manual describes how the user has to react in case of an error message so that the application can be continued if possible.Since there are several causes of errors that result in an error head (error head) message, no final root cause could be determined.It is noticeable that the evaluated error is largely due to a user error.Warning in the ifu regarding "hotplug" and label with the "hotplug" warning on the rfc have already been implemented in the past.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key8523820
MDR Text Key142645498
Report Number8010762-2019-00100
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,user facilit
Type of Report Initial,Followup
Report Date 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW DRIVE,BLUE
Device Catalogue Number701022161
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/03/2019
Initial Date FDA Received04/17/2019
Supplement Dates Manufacturer Received06/29/2019
Supplement Dates FDA Received07/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-