Model Number 10594738 |
Device Problems
Use of Device Problem (1670); Device Handling Problem (3265)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/29/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The multiple collision marks are not a result of regular system operation but improper use.The operator was advised by siemens not to use the concerned system in existing condition without repairs.
|
|
Event Description
|
Siemens service organization provided information about a mobile x-ray system - mobilett mira max - that represents risk in case of further clinical operation.The fork holding the x-ray tube assembly was found to be cracked and may potentially fall down.In addition siemens service organization found multiple visible collision marks on the machine.The user was advised not to use the unit until a respective correction is performed.There are no injuries attributed to this event.
|
|
Manufacturer Narrative
|
**resubmission of initial report as per fda on 04/03/2019** the investigation of the reported incident was completed with the following results.The fork and the single tank unit were returned for analysis.The single tank unit showed several collision marks; this indicates high mechanical impact on the tube assembly that led to the described damage of the fork.When driving the system the operator must avoid any collisions of the tube assembly to prevent system damage.The correct handling when moving the system is described in the operator manual (see xpr8-270.620.01.01.02 chapter "safety" page 22).The fork that holds the single tank is designed and tested to fulfill a four-fold mechanical safety.No general problem was identified.The spare part consumption of the fork shows values below the defined threshold.At the concerned customer site the issue was resolved by replacing the fork and the single tank unit.The system is fully functional.
|
|
Search Alerts/Recalls
|