The anesthesia workstation was investigated by the hospital biomedical engineer.At the start of investigation/test, the measured fico2 was about 10 mmhg.30 minutes later, the measured fico2 was close to zero.No parts were replaced.Evaluation of the received device logs can confirm the reported fico2 high alarms.The logs show further that the co2 absorber was bypassed and docked in place on three occasions and that the patient cassette was disconnected and replaced on two occasions, most likely during attempts to solve the high fico2 readings.The water trap shall be replaced at least every 30 days as stated in the user's manual.The filter in the water trap show a degradation in hydrophobic properties over time when it is used.This will be indicated by a larger pressure drop at moist conditions, and effects the co2 reading.It is unclear how long time the water trap had been in use at the time of the event.The root cause to the reported event has not been determined.
|