On (b)(6) 2019, a 27mm masters mechanical valve was selected for implant.After the aortic cross-clamp was removed, the left ventricular wall was confirmed to be bleeding.The tear was located near the native annulus, about 5-10mm below the native annulus.The tear was surgically closed.A small perforation was also observed.The user reported that the bleeding could be due to implant of too large size of the valve, or maybe caused by unconscious damage during the procedure.The 27mm was explanted and a 25mm masters was implanted.The patient is reported to be recovering.
|
An event of a tear was reported.The investigation confirmed the device met visual specifications when analyzed at abbott, with no damage or anomalies seen to the valve.The valve rotated and the leaflets opened and closed without resistance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the tear could not be conclusively determined; however, a smaller, 25mm, valve was implanted and per the site "the bleeding could be due to implant of too large size of the valve, or maybe caused by unconscious damage during the procedure".
|