• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS INTELLIVUE INFORMATION CENTER IX; CENTRAL STATION MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS INTELLIVUE INFORMATION CENTER IX; CENTRAL STATION MONITOR Back to Search Results
Model Number 866389
Device Problem Device Alarm System (1012)
Patient Problem Ventricular Tachycardia (2132)
Event Date 04/13/2019
Event Type  Injury  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer reported that an alarm did not go off on a piic ix when a patient experienced vtach rhythm on (b)(6) 2019.
 
Manufacturer Narrative
A philips field service engineer (fse) went to the customer site.It was determined that there was no product malfunction as the user had silenced alarms, just prior to the expected vtach alarm.It was also noted that the default profile appeared to have been changed.This information was provided to the customer.The device remains in use at the customer site.No subsequent calls have been logged for this device/issue.No further investigation or action is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE INFORMATION CENTER IX
Type of Device
CENTRAL STATION MONITOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
betty harris
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key8527380
MDR Text Key142359202
Report Number1218950-2019-02919
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/16/2019
Initial Date FDA Received04/18/2019
Supplement Dates Manufacturer Received04/16/2019
Supplement Dates FDA Received11/01/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-