• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems Disconnection (1171); Migration or Expulsion of Device (1395); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Burning Sensation (2146); Electric Shock (2554)
Event Date 04/07/2019
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 4351-35, serial# unknown, product type: lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: unknown, (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins) for gastrointestinal/pelvic floor issues.It was reported that one of the leads was no longer attached, which was diagnosed last week, and the patient was being ¿shocked¿ internally.It was noted that the doctor tried to shut the lead wire off, but the ¿zapping got worse¿, so they turned it back on, and since the day after turning stimulation back on, the patient was getting shocked every 3-5 seconds and it was worse when the patient was moving.The patient did not know how long they could handle the ¿burning sensation inside¿.The patient was redirected to a healthcare provider (hcp) regarding their symptoms.No further complications were reported/anticipated.
 
Manufacturer Narrative
Product id 4351-35, serial# unknown, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a healthcare professional (hcp) via a manufacturer representative.It was reported that a surgeon would be performing revision surgery on the patient in the coming weeks.No date was available at the time of the report.No further complications were reported/anticipated.
 
Manufacturer Narrative
Concomitant medical products: product id: 4351-35, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a healthcare professional (hcp) via a manufacturer representative.It was reported that a surgeon would be performing revision surgery on the patient in the coming weeks.No date was available at the time of the report.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8537763
MDR Text Key142784842
Report Number3004209178-2019-08002
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2019
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/18/2019
Initial Date FDA Received04/22/2019
Supplement Dates Manufacturer Received04/17/2019
05/13/2019
Supplement Dates FDA Received05/09/2019
05/20/2019
Date Device Manufactured12/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age44 YR
-
-