DEPUY ORTHOPAEDICS INC US SUMMIT POR TAPER SZ6 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1570-11-120
Device Problems Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Ossification (1428); Adhesion(s) (1695); Cyst(s) (1800); Foreign Body Reaction (1868); Unspecified Infection (1930); Muscular Rigidity (1968); Scar Tissue (2060); Tissue Damage (2104); Discomfort (2330); Test Result (2695); Muscle/Tendon Damage (4532); Unspecified Musculoskeletal problem (4535); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/13/2011
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Ppf alleges infection, metal wear, metallosis and elevated metal ions.Doi: (b)(6) 2008; dor: (b)(6) 2011; (right hip) first revision.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Updated code for metallosis.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  h10 additional narrative:   udi: (b)(4).Corrected: h6 (patient).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h6 clinical code: appropriate term / code not available (e2402) is used to capture blood heavy metal increased.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: g1.
 
Event Description
In addition to what previously alleged, pfs alleges stiffness and discomfort.After review of medical records patient was revised was due to loose acetabular component.Operative notes indicated that there were osteophytes present superiorly and anteriorly as well as soft tissue had to be debrided.The broken screw could be seen at that point.It appeared that the flange on the screw head had broken off or worn away and the screw had bent with almost 90 degrees angle posteriorly as the cup traveled posteriorly.The overall quality of the acetabulum was good.There were cystic areas but there were also very sclerotic areas an unusual undulating areas of sclerotic bone presumably due to metal impingement and reaction to wear.There were adhesion present, tendonitis, some slightly grayish fluid and the screw was sitting proud.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT POR TAPER SZ6 HI OFF
Type of Device
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key8543761
MDR Text Key142877831
Report Number1818910-2019-91401
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295060079
UDI-Public10603295060079
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1570-11-120
Device Catalogue Number157011120
Device Lot NumberCP8E71000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/02/2019
Initial Date FDA Received04/24/2019
Supplement Dates Manufacturer Received09/26/2019
11/01/2019
02/25/2020
11/16/2021
02/11/2022
Supplement Dates FDA Received10/14/2019
11/11/2019
02/25/2020
11/30/2021
02/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/25/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
12/14 ARTICUL 40MM M SPEC+5; PINN CAN BONE SCREW 6.5MMX30MM; PINN SECTOR W/GRIPTION 56MM; PINNACLE MTL INS NEUT40IDX56OD; SUMMIT POR TAPER SZ6 HI OFF; UNKNOWN HIP IMPLANT
Patient Age44 YR
Patient SexMale
Patient Weight114 KG
-
-