• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 44MM M SPEC+1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 44MM M SPEC+1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 136561000
Device Problems Naturally Worn (2988); Noise, Audible (3273)
Patient Problems Cyst(s) (1800); Pain (1994); Tissue Damage (2104); Weakness (2145); Hypoesthesia (2352); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Event Date 09/18/2018
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Claim and medical records received.After review of claim and medical records, it was stated that the patient was revised to address pain and aseptic loosening of cup component.Stiffness, limping, clicking, popping, weakness and numbness were also noted.Operative notes reported slightly brown tented and translucent fluid, the tissue is stained dark gray consistent of metallosis, two superior cysts were curetted and there was no acetabular bony ingrowth noted.When removing the acetabular shell, three screws broke but only one was fully extracted and the two screws were recessed since they were embedded into the sclerotic bone.Mri reported trochanteric bursitis.X-ray reported aseptic loosening of acetabular component and has metal wear debris reaction.Doi: (b)(6) 2010; dor: (b)(6) 2018; left hip.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 44MM M SPEC+1.5
Type of Device
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8544307
MDR Text Key142902642
Report Number1818910-2019-91467
Device Sequence Number1
Product Code JDI
UDI-Device Identifier10603295034032
UDI-Public10603295034032
Combination Product (y/n)N
PMA/PMN Number
K060031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2012
Device Catalogue Number136561000
Device Lot Number2360843
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/02/2019
Initial Date FDA Received04/24/2019
Supplement Dates Manufacturer Received05/09/2019
05/30/2019
Supplement Dates FDA Received05/09/2019
05/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient Weight98
-
-