• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION COMPONENT; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION COMPONENT; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Catalog Number EC-05500
Device Problem Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that an arrow epidural catheter could not be removed from the patient because it was splitting and shearing.The patient was transferred to another facility to have the catheter surgically removed.The catheter was surgically removed intact.
 
Manufacturer Narrative
Qn#(b)(4).No lot number was provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no relevant findings.Visual inspection could not be performed as no sample was returned by the customer for investigation.The customer did provide a photo that shows a piece of a catheter.The ifu for this kit, e-17019-110a; rev.2, was reviewed as a part of this complaint investigation.The ifu warns the user, "never tug or quickly pull on catheter during removal from patient to reduce risk of catheter breakage.Do not apply additional tension on the catheter if catheter begins to stretch excessively.Reposition patient to open the vertebral interspaces and re-attempt removal if resistance is encountered or if catheter stretches excessively during removal." a corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related cause.The customer did provide a photo that shows a catheter piece.However, without the actual sample, the potential cause of the catheter being difficult to remove could not be determined based upon the information provided.
 
Event Description
It was reported that an arrow epidural catheter could not be removed from the patient because it was splitting and shearing.The patient was transferred to another facility to have the catheter surgically removed.The catheter was surgically removed intact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION COMPONENT
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8549411
MDR Text Key143059948
Report Number1036844-2019-00419
Device Sequence Number1
Product Code BSO
Combination Product (y/n)N
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEC-05500
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/05/2019
Initial Date FDA Received04/25/2019
Supplement Dates Manufacturer Received05/21/2019
Supplement Dates FDA Received05/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
-
-