• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA VIEWFLEX¿ XTRA ICE CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA VIEWFLEX¿ XTRA ICE CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number D087031
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/08/2019
Event Type  malfunction  
Manufacturer Narrative
One viewflex xtra ice catheter was received for complaint evaluation.Visual inspection revealed a fracture in the shaft 0.5¿ proximal to the distal tip with exposed wires.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the fracture is due to damage during use.
 
Event Description
This report is to advise of an event observed during analysis confirming fracture in the shaft of the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIEWFLEX¿ XTRA ICE CATHETER
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8550958
MDR Text Key143421455
Report Number3008452825-2019-00215
Device Sequence Number1
Product Code OBJ
UDI-Device Identifier05415067002082
UDI-Public05415067002082
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K133853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model NumberD087031
Device Lot Number6687696
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/24/2019
Initial Date FDA Received04/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-