It was reported that a server room at the customer's facility had a power outage in october 2016, which resulted in the failure of the raid (hard disk unit) attached to the siemens syngo imaging system.A previous power outage that occurred earlier (september 2016) had already deactivated the raid mirroring function.Therefore, the data was not protected by any back-up option.After the assessment was conducted, it was determined that 15 potentially clinically relevant studies dated between october 13th and 15th were lost.This data will not be available for diagnosis if required by physician.The syngo imaging system did not contribute to the reported data loss.No consequences have been reported from this customer.The reported event occurred in (b)(6).
|