• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Entrapment of Device (1212)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2019
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformances related to the reported event were noted.A getinge field service engineer was dispatched to investigate.The fse evaluated the iabp unit and was able to duplicate the reported issue and found that the retraction issue had caused damage to the outer skin of the power cord.The fse confirmed that the iabp unit could plug into ac power and unlocked the reel and replumbed the power cord.All tests were performed and passed to factory specifications and the iabp unit was returned to the customer f and cleared for clinical service.
 
Event Description
It was reported that prior to use on a patient while preparing the cardiosave intra-aortic balloon pump (iabp) for use, the power cord was locked in the reel and could only be pulled out 15cm.The power cord could not be stretched, was unable to retract and could not be connected to ac power.The customer had to use a patch panel connection to connect the iabp unit to ac power.There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key8557188
MDR Text Key143857282
Report Number2249723-2019-00683
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-45
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Device AgeYR
Initial Date Manufacturer Received 04/03/2019
Initial Date FDA Received04/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-