The reported event could only partially be confirmed, since the device was returned, but no x-rays were provided.More information such as when the device was implanted, the patient's data and activity level, x-rays.Etc, are needed in order to determine the exact root cause of the complaint.The device inspection revealed the following: the product was returned and shows slight signs of wear such as deformed edges of the material, and loss of material in some areas.However, the device is not cut in two.Unfortunately, the analysis of the returned device alone is not sufficient to draw conclusions on this case.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|