• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH SYNGO.PLAZA; PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH SYNGO.PLAZA; PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM Back to Search Results
Model Number 10863171
Device Problems Use of Device Problem (1670); Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/22/2017
Event Type  malfunction  
Manufacturer Narrative
Resubmission of initial report as per fda on 4/3/2019.Activities are ongoing to evaluate the exact amount of affected data; however, the affected data is considered as clinically relevant.The product syngo.Plaza did not contribute to the data loss, as the issue is related to a hardware failure and the syngo.Plaza is a software solution only (b)(6).
 
Event Description
Per the potential complaint report, there was no injury associated with this issue.Due to an identified hardware failure in network attached storage (nas), there are missing folders containing patient images in the configured nas (nas1).A backup nas is configured, however, it is managed by a 3rd party application and not configured in the syngo.Plaza system.Contents of nas1 and nas2 are identical.Both nas systems have a large amount of empty folders for dates ranging between january 6, 2017 and april 19, 2017.The reported issue occurred in (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNGO.PLAZA
Type of Device
PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
henkestrasse 127
erlangen, 91052
GM  91052
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
henkestrasse 127
erlangen, 91052
GM   91052
Manufacturer Contact
anastasia sokolova
40 liberty blvd, mc 65-1a
malvern, PA 19355
6104486478
MDR Report Key8560710
MDR Text Key162081839
Report Number3002808157-2018-11604
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K132532
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number10863171
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/22/2017
Initial Date FDA Received04/29/2019
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-