• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. REFOBACIN REVISION 1X40G; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. REFOBACIN REVISION 1X40G; BONE CEMENT, ANTIBIOTIC Back to Search Results
Model Number 4011630001
Device Problem Improper Chemical Reaction (2952)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - event occurred in (b)(6).The device manufacturing quality record review did not show any non-conformity or deviation related to the described event.A sample of the same lot number has been analyzed in the laboratory under standardized conditions.The product did not show any unusual behavior during mixing, handling or setting.The reported behavior of the cement could not be reproduced.With the available information, the exact root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
 
Event Description
It has been reported that the cement was not solidified for 30 minutes during the surgery.The event occured on two cements of same reference and same batch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFOBACIN REVISION 1X40G
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e, bell dr.
warsaw, IN 46582
5745273773
MDR Report Key8563987
MDR Text Key143573804
Report Number3006946279-2019-00247
Device Sequence Number1
Product Code MBB
UDI-Device Identifier07350023774307
UDI-Public(01)07350023774307
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K150850
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Model Number4011630001
Device Catalogue Number4011630001
Device Lot NumberA743DC2204
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/03/2019
Initial Date FDA Received04/30/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-