Catalog Number 306575 |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Information (3190)
|
Event Date 04/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: unknown.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.Device manufacture date: unknown.
|
|
Event Description
|
It was reported that before use of the bd posiflush¿ sp pre-filled flush syringe nacl 0.9% there was no labeling.The following information was provided by the initial reporter: no labeling.
|
|
Event Description
|
It was reported that before use of the bd posiflush¿ sp pre-filled flush syringe nacl 0.9% there was no labeling.The following information was provided by the initial reporter: no labeling.
|
|
Manufacturer Narrative
|
Investigation: bd has been provided with photos and a sample for catalog 306575 to investigate for this record.One sample without it's flow warp was received.The sample had a scratch at the syringe barrel and does not have label attached.As a result, bd was able to verify the reported issue based on the photos provided.Syringes with no label are typically rejected by the machine.In this case, due to a punctual failure, the syringe could have been stuck on the rejection stations and finally went to the good rail and packaged.
|
|