EDWARDS LIFESCIENCES, PR FOGARTY ARTERIAL EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 120804F
Device Problems Break (1069); Device Damaged by Another Device (2915)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2019
Event Type  malfunction  
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product, a supplemental report will be sent with the investigation results.A device history record review was completed and documented that the device met all specifications upon distribution.
 
Event Description
It was reported that during use in surgery, the balloon on a fogarty catheter was punctured by the broken catheter tip.The issue was noted when the catheter was pulled out of the vessel when retrieving a clot.There were no complications and no additional interventions were needed.There was no allegation of patient injury.Patient demographics were requested but are not available.
 
Manufacturer Narrative
Our product evaluation laboratory received one model 120804f catheter.As received, a piece of the catheter body measuring approximately 5mm, was broken approximately 7mm from the catheter tip.The catheter body edges at the damaged site appeared to match up.The latex balloon was pierced by the broken catheter body part.The latex edges at the puncture site appeared to match up.Both windings were intact.The lab findings appeared consistent with the customer photo.No other visible damage was observed from the catheter.The customer report of "the balloon was punctured with the broken catheter tip" was confirmed on evaluation.An engineering evaluation has been initiated to assess for any manufacturing-related processes which could be correlated to the complaint.Balloon rupture and catheter separation, as a result of excessive pull force applied to remove adherent material, are the most frequent causes of reported failures.To minimize the risk of vessel damage, balloon rupture, or tip detachment, the maximum recommended inflation volume and pull force for the catheter should not be exceeded.The arterial embolectomy catheter is not recommended for the removal of fibrous, adherent, or calcified material (e.G.Chronic clot, atherosclerotic plaque).This catheter is not designed to withstand the additional pull force needed to remove these materials.The instructions for use of the product contains the following statement: ¿as with all catheterization procedures, complications may occur.These may include local or systemic infection, local hematomas, intimal disruption, arterial dissection, perforation and vessel rupture, hemorrhage, arterial thrombosis, distal embolization of blood clots and atherosclerotic plaque, air embolus, aneurysm, arterial spasm, arteriovenous fistula formation, and balloon rupture with fragmentation, tip separation and distal embolization.¿ complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY ARTERIAL EMBOLECTOMY CATHETER
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key8570729
MDR Text Key143975423
Report Number2015691-2019-01555
Device Sequence Number1
Product Code DXE
Combination Product (y/n)N
PMA/PMN Number
PREAMENDMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/18/2020
Device Model Number120804F
Device Catalogue Number120804F
Device Lot Number61516409
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2019
Initial Date Manufacturer Received 04/11/2019
Initial Date FDA Received05/01/2019
Supplement Dates Manufacturer Received05/17/2019
07/23/2020
Supplement Dates FDA Received05/30/2019
01/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-