• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM-CONTROLLER 2.0; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM-CONTROLLER 2.0; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1420
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/26/2019
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Concomitant medical products: 1103 vad, implanted on: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the controller exhibited a malfunctioning display.The controller was exchanged.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Correction: d4 product event summary: the controller was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the controller revealed that the unit passed functional testing.A visual inspection under 10x magnification revealed hairline cracks around both power ports one and two (1<(>&<)>2).An internal inspection did not reveal evidence of fluid ingress.The observed hairline cracks are not related to the reported event.Based on an investigated, the root cause of the hairline crack(s) was determined to be due to chemical additives applied to the power port gaskets during the manufacturing process.The chemical additives contributed to environmental stress cracking.Failure analysis of the controller revealed that the controller was able to display all characters properly.As a result, the reported "malfunctioning display" event could not be confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM-CONTROLLER 2.0
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key8573837
MDR Text Key143845039
Report Number3007042319-2019-06870
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707000475
UDI-Public00888707000475
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2019
Device Model Number1420
Device Catalogue Number1420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2019
Initial Date Manufacturer Received 04/29/2019
Initial Date FDA Received05/02/2019
Supplement Dates Manufacturer Received06/24/2019
Supplement Dates FDA Received06/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-0067-2019
Patient Sequence Number1
Treatment
CD3369-40Q ICD, 439888 LEAD; CD3369-40Q ICD, 439888 LEAD
Patient Age37 YR
-
-