• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problem Increase in Pressure (1491)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 03may2019.A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer reported an over pressure condition at turn on.There was no patient involvement.Event date not specified: estimate used.
 
Manufacturer Narrative
Date of report: 07jun2019.Date received by manufacturer: 09may2019.The customer confirmed that reported flow valve issue.The customer replaced the flow valve assembly.Customer responded via email that all testing passed.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8578245
MDR Text Key144138850
Report Number2031642-2019-02761
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/15/2019
Initial Date FDA Received05/03/2019
Supplement Dates Manufacturer Received04/15/2019
Supplement Dates FDA Received06/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-