The user facility requested the testing to determine if the table caused a patient to slip off the table.No report of injury.A procedure delay was reported.A steris service technician arrived onsite to inspect the 5085 surgical table and found the table to be operating properly.The table performed as commanded; no issue with the function or operation of the table was identified and the table was returned to service.The user facility reported that during the time of the reported event, the table was tilted head down and left with a safety strap across the patient's lap when the patient slipped of the table.The reported event is attributed to user error as user facility personnel should have ensured the patient was properly secured.The 5085 surgical table operator manual states (1-2), "warning - personal injury hazard: failure to keep the patient properly secured with the patient safety straps at all times could result in death or serious injury." the operator manual further states (1-2), "warning - personal injury hazard: unanticipated table movement could cause patient injury.Patient must be secured to the table in accordance with recommended positioning practices." a steris account manager counseled user facility personnel on ensuring the patient was properly secured.No additional issues have been reported.
|