• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: POSEY PRODUCTS LLC POSEY BED 8070; PATIENT BED WITH CANOPY/RESTRAINTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

POSEY PRODUCTS LLC POSEY BED 8070; PATIENT BED WITH CANOPY/RESTRAINTS Back to Search Results
Model Number 8070
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Customer confirmed will not be returning product.Without the device the reported issue could not be confirmed and possible causes of the failure could not be determined.This product is a serviceable item and historical review of similar complaints against this product was determined to be related to normal wear and tear.At this time, there is no evidence that a manufacturing non-conformity contributed to the reported complaint and the instructions for use (ifu) were reviewed and determined to provide adequate instructions and warning for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warrant.Manufacturer reference file # (b)(4).Product will not be returning.
 
Event Description
The patient's guardian reported the canopy is ripping at the seams inside and outside.The netting has holes and the front of the canopy has been glued by the guardian.The canopy is in very bad condition.Additional information from the guardian, received on 2019/04/12, stated the issues with the canopy was discovered when the user fell through the canopy due to the split in the seams.Guardian confirmed the patient did not suffer any injury from the fall.The date the issue was discovered is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8070
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
POSEY PRODUCTS LLC
5635 peck rd
arcadia CA 91006
Manufacturer Contact
william hincy
posey company
5635 peck road
arcadia, CA 91006
6264433143
MDR Report Key8579234
MDR Text Key145506296
Report Number2020362-2019-00123
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113357
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 04/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number8070
Device Catalogue Number8070
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/12/2019
Initial Date FDA Received05/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-