MAUDE Adverse Event Report: MMJ SA DE CV (USD) MON-A-THERM; THERMOMETER, ELECTRONIC, CLINICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) MON-A-THERM; THERMOMETER, ELECTRONIC, CLINICAL Back to Search Results
Model Number 90050
Device Problem Low Readings (2460)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/12/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the device temperature had a low display.It was reported that the device was replaced.There was no patient harm.
 
Manufacturer Narrative
Additional information: d10, g4, h3, h6 h3.Evaluation code: one sample of temperature probe was received for evaluation, no lot number was provided, a visual inspection was performed and it was observed all the components are assembled properly at the probe.A functional test was performed on the thermistor, using the liquid bath reading of the temperature was taken from the digital thermometer calibration, the thermistor was connected to the digital multimeter with calibration to test its resistance while in the bath.Therefore, no failure mode was observed in the received sample since met with the product specifications and no corrective actions are required.The device history record (dhr) was not reviewed since the lot number was not provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MON-A-THERM
Type of Device
THERMOMETER, ELECTRONIC, CLINICAL
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
MDR Report Key8583150
MDR Text Key144163149
Report Number2936999-2019-00327
Device Sequence Number1
Product Code FLL
UDI-Device Identifier20884521031606
UDI-Public20884521031606
Combination Product (y/n)N
PMA/PMN Number
K874514
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 09/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number90050
Device Catalogue Number90050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2019
Initial Date Manufacturer Received 04/16/2019
Initial Date FDA Received05/06/2019
Supplement Dates Manufacturer Received08/15/2019
Supplement Dates FDA Received09/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-