• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Physical Resistance/Sticking (4012)
Patient Problem No Patient Involvement (2645)
Event Date 04/15/2019
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint for the platform displayed an error message user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) was confirmed during archive data review and functional testing.Load cell was found defective and a replacement was required.In addition, the corrosion was present on the drive train motor and the motor drive train brake completely locked due to the built rust, thus confirming the customer reported complaint of the lifeband unable to retract.Visual inspection was performed and found extensive corrosion and fluid damage inside the top cover, and front and bottom enclosure, unrelated to the reported issue.Per autopulse user guide, 4.2.Cleaning the autopulse platform section: "wipe all the surfaces of the autopulse platform free of foreign matter and spills with a disinfectant or bactericidal wipe.Check the vents to ensure that they are free and clear of any obstructive matter.Caution: do not submerge the autopulse in liquid.Ensure that the autopulse is dry before storing." during functional testing, user advisory (ua) 07 error message displayed upon power on the device, thus confirming the reported complaint.The load characterization check was performed and verified that one of the load cells is out of specification.Review of the archive data indicated multiple error messages user advisory (ua) 07 around the customer reported event date.Due to the received condition of the device, no further service will be performed on the platform.The device will be returned to the customer labeled "not for clinical use".Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of reported complaints for autopulse platform sn (b)(4).
 
Event Description
During shift check, the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message and the lifeband would not retract.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
noemi schambach
2000 ringwood ave,
san jose, CA 95131
4084192955
MDR Report Key8589270
MDR Text Key144607573
Report Number3010617000-2019-00371
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/15/2019
Initial Date FDA Received05/07/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-