DEPUY SYNTHES PRODUCTS LLC BLACKMAX-NEURO LOW PRESSURE; MOTOR, DRILL, PNEUMATIC - HANDPIECE Back to Search Results
Catalog Number BLACKMAX-N-LP
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/16/2019
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use over time.If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi ¿ (b)(4).
 
Event Description
It was reported from (b)(6) that during service and evaluation, it was determined that the handpiece device had a high-pressure leak.It was noted that the handpiece device was making noise and leaking air.It was further determined that the device failed pretest for noise verification.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLACKMAX-NEURO LOW PRESSURE
Type of Device
MOTOR, DRILL, PNEUMATIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6103142063
MDR Report Key8591141
MDR Text Key144686039
Report Number1045834-2019-53560
Device Sequence Number1
Product Code HBB
UDI-Device Identifier00845384001102
UDI-Public845384001102
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K831756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBLACKMAX-N-LP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2019
Initial Date Manufacturer Received 05/03/2019
Initial Date FDA Received05/08/2019
Date Device Manufactured10/03/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-