DEPUY SYNTHES PRODUCTS LLC BLACKMAX-NEURO LOW PRESSURE; MOTOR, DRILL, PNEUMATIC - HANDPIECE
|
Back to Search Results |
|
Catalog Number BLACKMAX-N-LP |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use over time.If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi ¿ (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the handpiece device had a high-pressure leak.It was noted that the handpiece device was making noise and leaking air.It was further determined that the device failed pretest for noise verification.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|