• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems High impedance (1291); Unintended Collision (1429)
Patient Problems Edema (1820); Nausea (1970)
Event Date 04/16/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) regarding a patient with an implantable neurostimulator (ins).It was reported that there was edema or fluid in the ins pocket site per a ct scan.The patient was implanted on (b)(6) 2019 and was reporting nausea.The hcp reported impedances c2-255 ohms and c3 ¿ 901 ohms (>800) and c2 < c3 1011 ohms.The hcp wondered if there was an open circuit.They indicated a ct was performed and it looked like there was fluid in the pocket.Additional information from a manufacturer representative (rep) stated the hcp thought it was post-implant edema.No further patient complications were reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from a manufacturer representative (rep).It was reported that the patient experienced a fall on their abdomen a few weeks prior to the report.The hcp believed the fall may have put pressure on one of the leads, resulting in a 901 [ohm] impedance reading.The hcp placed the patient in unipolar stimulation mode, using c2 as the stimulating electrode, turning off lead c3, and making case positive.They were looking to get the patient scheduled for the operating room (or) in a couple of weeks to open the pocket site and tighten the setscrews to normalize the impedance.Clinically and symptomatically, the patient received significant benefit from the device and continued to do well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
MDR Report Key8591362
MDR Text Key145379380
Report Number3007566237-2019-01042
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/06/2019
Initial Date FDA Received05/08/2019
Supplement Dates Manufacturer Received05/08/2019
Supplement Dates FDA Received05/23/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age62 YR
-
-