• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR); PACEMAKER PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR); PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/29/2019
Event Type  malfunction  
Event Description
It was reported that while updating the software on the programmer in clinic, a burning smell was noted and the screen went black.Nothing was blocking the fan.The programmer was replaced.No patient was involved.
 
Manufacturer Narrative
The reported event of a dim screen and burning smell was verified.During analysis, the unit was plugged in to a working outlet and power on function was performed.The unit powered up, but the display was dim.Suspect components were analyzed and revealed that the inverter board suffered from a failure where one of it's components were burned.The failure was contained within the housing and posed no risk of exposure to the user or patient.Inverter board failure is the cause of complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8592256
MDR Text Key144461028
Report Number2017865-2019-06646
Device Sequence Number1
Product Code KRG
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/29/2019
Initial Date FDA Received05/08/2019
Supplement Dates Manufacturer Received06/14/2019
Supplement Dates FDA Received06/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-