ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP-JAPAN
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A product sample was received at the manufacturing site and is awaiting evaluation.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported there was leakage from the cassette and it failed testing during priming.There was no patient involvement.The procedure was performed and completed after replacing the product with another one.
 
Manufacturer Narrative
The lot complaint history was reviewed, this is the tenth complaint for the finish goods lot; however, the first for this issue for this lot.The device history record shows the product was released per specifications.The returned sample was visually inspected and there were no obvious defects observed that would have contributed to the reported event.A full product pressure leak test was then conducted on the cassette utilizing an external pressure source and no leakage was detected.A calibrated console representing a current software version was then used to test the sample.The sample could prime, and pass intraocular pressure (iop) calibration successfully.No anomalies were observed during priming.No system message code was generated during testing.No leakage was detected from the manifolds, connectors, or drain path between the consumable and console.After functional testing no leakage was detected from the pump elastomer or on the pump area of the fluidics module.The presence of fluid leaking from the cassette was not confirmed.After both leakage and console laboratory testing, inspection of the sample indicated no signs of continued leakage from the pump elastomer or cassette.Furthermore, there were no signs of fluid leakage onto the fluidics console due to leakage from the cassette.The cassette passed functional and performance testing.This complaint has been reviewed and the sample was found to be conforming, therefore no further actions will be pursued at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8596444
MDR Text Key144689624
Report Number2028159-2019-00854
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-JAPAN
Device Catalogue Number8065751817
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2019
Initial Date Manufacturer Received 04/22/2019
Initial Date FDA Received05/09/2019
Supplement Dates Manufacturer Received07/19/2019
Supplement Dates FDA Received08/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION SURGICAL PROCEDURE PAK
-
-