MAUDE Adverse Event Report: BATTLE CREEK EQUIPMENT THERMOPHORE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BATTLE CREEK EQUIPMENT THERMOPHORE Back to Search Results
Model Number 155
Device Problems Thermal Decomposition of Device (1071); Smoking (1585); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Customer stated, "the pad was smoking and i saw it had burned through the cover" the product was returned for investigation.The customer did not claim any injury.An investigation was done to look into the customers complaint.The investigator observed that a thermostat had burned.This caused the burn mark the customer described.The thermostat burn was caused by the user misusing the pad by folding and laying on the pad while in use.Ifu states "do not sit on, lie on, or crush pad.Avoid sharp folds.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE
Type of Device
THERMOPHORE
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
702 s reed rd
fremont IN 46737
Manufacturer Contact
uday bahal
MDR Report Key8597621
MDR Text Key144808179
Report Number1832415-2019-10363
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 04/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number155
Device Lot Number00155194182
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/09/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-