|
Catalog Number 466P306 |
Device Problem
Failure to Align (2522)
|
Patient Problems
Perforation (2001); Perforation of Vessels (2135)
|
Event Date 10/30/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
As reported, the patient underwent placement of a trapease inferior vena cava (ivc) filter.The indication for filter placement is not available.The filter subsequently malfunctioned and caused injury and damage to the patient, including but not limited to perforation and tilt.The filter remains implanted; thus, unavailable for analysis.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without procedural films for review, the filter tilt reported could not be confirmed.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
|
Event Description
|
As reported by the legal brief, the patient underwent placement of a trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damage to the patient, including but not limited to perforation and tilt.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
|
|
Manufacturer Narrative
|
Complaint conclusion: as reported, the patient underwent placement of a trapease vena cava filter.The patient is reported to have had a history of deep vein thrombosis (dvt) and pulmonary embolism (pe) with a contraindication to anticoagulation.The filter was implanted via the right common femoral vein and deployed with good alignment with its¿ tip at the l1-l2 junction.The patient is reported to have tolerated the procedure well.Approximately four years and two months after the implantation, the patient became aware that the filter had tilted and that the strut(s) had perforated outside the inferior vena cava (ivc).The patient further reported having experienced anxiety and anxiety associated with the filter.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The trapease vena cava filter is indicated for use in the prevention of recurrent pe via percutaneous placement in the ivc for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without images or procedural films for review, the reported filter tilt and ivc perforation events could not be confirmed and the exact cause could not be determined.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.Additionally, the timing and mechanism of the filter tilt is unknown.It is unknown if the tilt contributed to the reported perforation.A review of the instructions for use (ifu) notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include the patient¿s pre-existing co-morbidities, pharmacological issues and lesion characteristics.The anxiety experienced by the patient does not represent a device malfunction.Anxiety, part of the body¿s natural response to stress and can cause feelings of, but not limited to, nervousness, unease and worry.Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
|
Event Description
|
The inferior vena cava filter was placed via the right common femoral vein approach.There was no sign of intracaval thrombus and the cava was of adequate caliber to receive the filter.The filter was placed with the tip at the l1-2 junction, it was aligned well.The patient tolerated the procedure without difficulty.
|
|
Search Alerts/Recalls
|
|
|