Model Number RVDR01 |
Device Problems
Incorrect Measurement (1383); Unable to Obtain Readings (1516); Loss of Data (2903); Missing Test Results (3267)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/16/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical product: 5076-52, lead, implanted: (b)(6) 2015.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the implantable pulse generator (ipg) battery voltage could not be determined.It was noted that a bit flip may have occurred.The ipg remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the battery measurement was not available.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that the patient underwent radiation therapy which affected the histograms, and the implantable pulse generator (ipg) was interrogated in order to resolve the issue of battery voltage data not being collected.
|
|