• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 2ACH20
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Great Vessel Perforation (2152)
Event Date 04/18/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the patient¿s pulmonary vein was perforated by the mapping catheter when it was pulled into the left atrium.No treatment or extended hospitalization was required.The case was completed with cryo.The patient was in good condition after the procedure.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed seventeen applications were performed with balloon catheter 2af284 with lot number 52407.A clinical issue, a perforation, was encountered during the procedure.In conclusion, the mapping catheter was not returned for investigation.There is no indication of relation of adverse event to the performance of the cryo device.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8605371
MDR Text Key144875516
Report Number2182208-2019-00882
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K162892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/28/2020
Device Model Number2ACH20
Device Catalogue Number2ACH20
Device Lot Number11081812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2019
Initial Date FDA Received05/13/2019
Supplement Dates Manufacturer Received05/21/2019
Supplement Dates FDA Received06/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-