Catalog Number 210712
Device Problem Migration or Expulsion of Device (1395)
Patient Problems Not Applicable (3189); No Code Available (3191)
Event Date 05/02/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.No nonconformances were identified for this part-lot number combination per qlik query executed 05/14/2019.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Event Description
It was reported by the sales rep via phone that during a labrum repair procedure, the patient's bone hole was prepared and drilled and while malleting the anchor into the bone, the lupine loop rapide anchor with orthocord slipped out of the patient's bone while removing the drill guide.The sales rep stated that the loop on the anchor was exposed and therefore it was unusable.The anchor was removed from the patient by pulling on the sutures with no debris left behind.The case was completed with another anchor in a new bone hole.There was a two minute surgical delay to remove the anchor.The anchor was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUPINE BR DS W/ORTHCRD
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8612640
MDR Text Key145138111
Report Number1221934-2019-57153
Device Sequence Number1
Product Code MAI
UDI-Device Identifier10886705001101
UDI-Public10886705001101
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial
Report Date 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Catalogue Number210712
Device Lot Number2L25190
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/02/2019
Initial Date FDA Received05/15/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-