• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® CRITICORE® MONITOR; CRITICORE MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® CRITICORE® MONITOR; CRITICORE MONITOR Back to Search Results
Model Number 000002N
Device Problems Break (1069); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the criticore monitor was not reading correctly.
 
Event Description
It was reported that the criticore monitor was not reading correctly.
 
Manufacturer Narrative
The reported problem was confirmed.The root cause for the reported issue could not be determined; however, the device is a portable unit and is susceptible to having damage done to the display, enclosure, handles and arm assemblies through using and positioning the device.The monitor was given a general functionality test.This monitor was tested for functionality per the criticore analog series final system tests.The temperature sensor was calibrated and is functioning properly.The temperature probe jack had broken free of the display pcb (one o-ring was missing) and the unit cannot measure the temperature.It is probable that the jack was in the process of breaking off when the temperature was not reading correctly.The temperature probe jack was replaced.The investigation indicated that the reported issue was not manufacturing or supplier related.Therefore, a device history record review was not required.The instructions for use were found adequate and state the following: "warning: the criticore monitor requires special precautions regarding emc and needs to be installed and put into service according to the emc information provided in the following tables.Portable and mobile rf communications equipment may affect the criticore monitor.Warning: the criticore monitor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the criticore monitor should be observed to verify normal operation in the configuration in which it will be used.Warning: do not immerse or submerge the monitor or turn it upside down when cleaning.Caution: use only heavy-duty alkaline d cell batteries.Do not use rechargeable batteries.Do not incinerate batteries.Recycle or dispose of them properly.Contact bard for disposal information.Caution: improper orientation of the batteries within the battery pack can potentially damage the criticore monitor.Caution: state and federal regulations govern the packaging necessary for return of medical product which may have been contaminated.Refer to local, state and federal regulations when packaging the criticore monitor for return.Caution: there are no user serviceable components inside the criticore monitor.The user should not attempt to repair the criticore monitor.To do so may void the warranty and could result in erroneous monitor readings.Caution: use of cables or sensors other than those specified for use with the criticore monitor, except those sold by bard for use as replacement part or repair components, may result in increased emissions or decreased immunity of the criticore monitor.Caution: the criticore monitor should be recycled properly per european union directive 2002/96/ec on waste electronic and electrical equipment, january 27, 2003.Do not dispose with ordinary municipal waste." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® CRITICORE® MONITOR
Type of Device
CRITICORE MONITOR
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8616220
MDR Text Key145356898
Report Number1018233-2019-02611
Device Sequence Number1
Product Code EXS
UDI-Device Identifier00801741080777
UDI-Public(01)00801741080777
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 08/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number000002N
Device Catalogue Number000002N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2019
Initial Date Manufacturer Received 04/24/2019
Initial Date FDA Received05/16/2019
Supplement Dates Manufacturer Received07/22/2019
Supplement Dates FDA Received08/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-