Catalog Number 15125 |
Device Problem
Deflation Problem (1149)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/21/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#: (b)(4).
|
|
Event Description
|
The complaint is reported as: "the facility reported no function of the check valve (no deflation attainable) after 7-time usages".The issue was detected prior to use on a patient during inspection/functional testing.
|
|
Event Description
|
The complaint is reported as: "the facility reported no function of the check valve (no deflation attainable) after 7-time usages." the issue was detected prior to use on a patient during inspection/functional testing.
|
|
Manufacturer Narrative
|
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the airway tube of the device was yellowish.No damage was observed on the device.The device was unable to be inflated and deflated via the check valve when tested.A device history record (dhr) review was performed and there were no issues found that could be related to the reported complaint.Based on the investigation performed, the reported complaint was confirmed.It is suspected that some very tiny debris/substances might have been brought into the check valve while reprocessing over the years (since february 2013) that led to the valve mechanism failure.
|
|
Search Alerts/Recalls
|