DEPUY ORTHOPAEDICS, INC. 1818910 PINN MAR NEUT 32IDX54OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121932054
Device Problems Material Separation (1562); Device Dislodged or Dislocated (2923)
Patient Problems Joint Dislocation (2374); No Code Available (3191)
Event Date 04/26/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Marathon liner 54 x 32mm was placed in a pinnacle 100 cup on (b)(6) 2014.Liner dislocated on sunday (b)(6) 2019.Patient brought to the operating room for head and liner exchange on friday (b)(6) 2019.The liner was replaced with a 54mm neutral altrx liner.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the acetabular liner does find evidence to support the reported disassociation event.Product error / contribution to the event is not identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: 481159.Device history review: no related deviations or anomalies were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 32IDX54OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8619329
MDR Text Key145359356
Report Number1818910-2019-93784
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295013914
UDI-Public10603295013914
Combination Product (y/n)N
PMA/PMN Number
K001534
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number121932054
Device Lot Number481159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2019
Initial Date Manufacturer Received 04/29/2019
Initial Date FDA Received05/17/2019
Supplement Dates Manufacturer Received05/24/2019
06/26/2019
Supplement Dates FDA Received05/27/2019
06/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
-
-