• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 466P306X
Device Problem Failure to Align (2522)
Patient Problem Perforation of Vessels (2135)
Event Date 04/19/2018
Event Type  Injury  
Manufacturer Narrative
As reported, the trapease inferior vena cava (ivc) filter malfunctioned and caused injury, damage, to the patient, including, but not limited to perforation and tilt.The indication for filter placement is not available.The filter remains implanted; thus, unavailable for analysis.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without procedural films for review, the filter tilt reported could not be confirmed.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal brief, the trapease filter subsequently malfunctioned and caused injury, damage, the patient, including, but not limited to perforation and tilt, that causes injury and damage to the patient.
 
Manufacturer Narrative
As reported, the trapease filter subsequently malfunctioned and caused injury, damage, the patient, including, but not limited to perforation and tilt.Per the medical records, history includes deep venous thrombosis (dvt) and pulmonary embolism (pe).During the implant procedure, the trapease filter was deployed at the level of l1-l2.Per the patient profile from (ppf), the patient reports perforation of filter strut(s) outside the ivc, and tilt.The patient also reports anxiety and fear.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without procedural films for review, the filter tilt reported could not be confirmed.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Anxiety does not represent a device malfunction and may be related to underlying patient related issues.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
466P306X
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, tipperary 0000
EI  0000
MDR Report Key8620673
MDR Text Key145404724
Report Number1016427-2019-02870
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number466P306X
Device Catalogue Number466P306X
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/19/2019
Initial Date FDA Received05/17/2019
Supplement Dates Manufacturer Received06/08/2019
Supplement Dates FDA Received07/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age54 YR
-
-