MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-87008
Device Problem Entrapment of Device (1212)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 05/14/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During an atrial tachycardia procedure, the catheter became lodged in the patient and could not be removed.Following mapping in the right atrium, fluoroscopy revealed the catheter would not move and could not be removed from the inferior cava.A tee did not determine where the catheter was located, and the patient was transferred to surgery to extract the catheter via the subclavian vein.The patient is recovering in the uci.
 
Event Description
It is possible that the device was trapped in fibers between the inferior vena cava and the lateral wall.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
MDR Report Key8623564
MDR Text Key145497367
Report Number2030404-2019-00035
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05414734309929
UDI-Public05414734309929
Combination Product (y/n)N
PMA/PMN Number
K101623
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberIBI-87008
Device Lot Number6602300
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/14/2019
Initial Date FDA Received05/20/2019
Supplement Dates Manufacturer Received05/22/2019
Supplement Dates FDA Received05/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age31 YR
Patient Weight68
-
-