Catalog Number 130344
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event Description
During incoming inspection, the distributor rejected this device, 130344, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.Due to the potential severity of a breach in sterility, this complaint meets the criteria for a reportable event.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Manufacturer Narrative
Received one 130344 in unopened original packaging.Lot number was verified.Performed a visual inspection of the device, there were no obvious signs of a breach.Performed a functional inspection, the devices were dye leak tested which indicated that the packaging did not have an insufficient heat seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.There have been two compalints for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 4 complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: contraindications: these devices should never be used when: there is visible evidence of damage to the exterior of the device such as cuts, punctures, nick abrasions, unusual lumps or significant discoloration.Inspection: these devices should be inspected before each use.Visually examine the devices for obvious physical damage and do not use if damage is found: cracked, broken or otherwise distorted plastic parts broken or significantly bent connector contacts this issue will continue to be monitored through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE FUNCTION HNDPC ABC ELEC
Type of Device
APPARATUS, ELECTROSURGICAL
Manufacturer (Section D)
CONSOLIDATED MEDICAL EQUIPMENT
alejandro dumas ave, 11321
complejo industrial chihuahua
chihuahua, 31136
MX  31136
MDR Report Key8623650
MDR Text Key145504616
Report Number3007305485-2019-00150
Device Sequence Number1
Product Code HAM
Combination Product (y/n)N
PMA/PMN Number
K871435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 07/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/19/2023
Device Catalogue Number130344
Device Lot Number201806204
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2019
Initial Date Manufacturer Received 04/25/2019
Initial Date FDA Received05/20/2019
Supplement Dates Manufacturer Received07/11/2019
Supplement Dates FDA Received07/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-