• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H
Device Problem Defective Device (2588)
Patient Problems Hearing Impairment (1881); Therapeutic Response, Decreased (2271)
Event Date 03/14/2019
Event Type  malfunction  
Event Description
The recipient reportedly elected explant surgery due to a defect in the electrode.The recipient presented with decreased performance and sound quality issues.The recipient's device was explanted.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient reportedly healed following surgery.The external visual inspection revealed the electrode was cut prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed cut electrode wires.This is believed to have occurred during revision surgery.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The failure of this device is attributed to excessive moisture through a gross leak at the case-band braze.This ultimately caused this device to cease functioning.This older device configuration is no longer manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key8624749
MDR Text Key145623144
Report Number3006556115-2019-00271
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/19/2002
Device Model NumberAB-5100H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/20/2019
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age20 YR
-
-