U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 23may2019.The manufacturer's field service engineer (fse) performed troubleshooting but was unable to duplicate the reported error; however, the fse did confirm the error through the device event logs.The fse replaced the central processing unit (cpu) to address the findings.The unit passed performance verification testing.
 
Event Description
It was reported that the unit declared an error ventilator restarted due to anomalies detected during operation.It is unknown if the device was in use at the time of the event; however, there was no patient harm reported.
 
Manufacturer Narrative
Date rec¿d by mfr : 17aug2019.The central processing unit (cpu) board was returned to the manufacturer for failure analysis.The cpu board revealed no signs of damage or contamination.During testing, the reported failure could not be duplicated and no failures were identified.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8636189
MDR Text Key146061256
Report Number2031642-2019-03184
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/06/2019
Initial Date FDA Received05/23/2019
Supplement Dates Manufacturer Received05/06/2019
Supplement Dates FDA Received08/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-