• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 220; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 220; JCX Back to Search Results
Catalog Number PMX220
Device Problem Suction Problem (2170)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/13/2019
Event Type  malfunction  
Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 220v (pump max).During the procedure, the physician completed the first pass using the pump max.While attempting to perform the second pass, no vacuum was observed and the pressure gauge needle did not move.Therefore, the pump max was disconnected and the procedure was completed using a manual aspiration with a syringe.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
Results: dried blood was present inside the vacuum inlet nozzle.Conclusions: evaluation of the returned penumbra system aspiration pump max 220v (pump max) revealed dried blood was inside the vacuum inlet nozzle.If the aspiration tubing is connected directly to the vacuum inlet nozzle instead of the canister supplied by penumbra, blood will likely be aspirated into the vacuum assembly.If fluid is aspirated into the vacuum assembly, the pump may not function properly.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 220
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8636690
MDR Text Key145949725
Report Number3005168196-2019-01037
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012698
UDI-Public00814548012698
Combination Product (y/n)Y
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 01/01/2005,04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX220
Device Lot NumberF33654-26
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2019
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 04/26/2019
Initial Date FDA Received05/23/2019
Supplement Dates Manufacturer Received06/03/2019
Supplement Dates FDA Received06/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age83 YR
-
-