• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 220; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 220; JCX Back to Search Results
Catalog Number PMX220
Device Problems Misconnection (1399); Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/23/2019
Event Type  malfunction  
Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 220v (pump max).During the procedure, the aspiration tubing (tubing) was accidently connected directly to the pump max rather than the canister.Consequently, blood was aspirated into the pump max and, therefore, it was removed.The procedure was completed using a new pump max.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
Results: dried blood was on the pump housing and inside the vacuum inlet.Conclusions: evaluation of the returned pump max confirmed blood was aspirated into the pump.If fluid enters the vacuum pump assembly, the pump may not function properly.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 220
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8636715
MDR Text Key145954980
Report Number3005168196-2019-01005
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012698
UDI-Public00814548012698
Combination Product (y/n)Y
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/01/2005,04/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX220
Device Lot NumberF43591-16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2019
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 04/23/2019
Initial Date FDA Received05/23/2019
Supplement Dates Manufacturer Received07/09/2019
Supplement Dates FDA Received07/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-