Model Number 310301
Device Problem Nonstandard Device (1420)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2019
Event Type  malfunction  
Event Description
During an aquablation procedure, e22 (r motion error) message occurred.Aquabeam handpiece was replaced and the procedure was completed with no patient injury reported.
 
Manufacturer Narrative
H.10.[x] additional manufacturer narrative the aquabeam handpiece was returned for investigation.The aquabeam handpiece was connected to an aquabeam system and a test aquablation treatment run was performed; a leak was observed.The handpiece was disconnected, the outer shell was removed, and the handpiece was connected again to the aquabeam system and a test treatment run was performed without the outer shell.A leak was observed at the p1080 probe tube to p1160 pump cartridge distal crimp laser weld.The weld leak resulted in saline leaking on the sensor board, which caused an r position lost error.The error is displayed to the user as an e22 error on the aquabeam cpu.The investigation confirmed the weld leak likely caused the reported e22 error.The reported event was confirmed during the investigation.Root cause investigation into this issue determined that the micro-crack that can develop in a weld joint within the handpiece, and cause the failure mode, is due to variability in a single pass welding process.The contributing factors to the variability were identified through a fish bone analysis and are listed below.1.Weld placement 2.Component gap 3.Weld design 4.Weld inspection 5.Material compatibility 6.Increase in production volume multiple long-term corrective and preventive actions were established to further address the underlying factors identified.This event is part of recall 3012977056-030619-001-r.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
H.11 corrected data: for corrected data, please refer to section h.1 type of reportable event, follow-up #1 was incorrectly submitted as "serious injury".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS
900 island drive, suite 101
redwood city 94065
MDR Report Key8638811
MDR Text Key146359430
Report Number3012977056-2019-00025
Device Sequence Number1
Product Code PZP
Combination Product (y/n)N
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Model Number310301
Device Lot Number18C00153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2019
Initial Date Manufacturer Received 04/23/2019
Initial Date FDA Received05/23/2019
Supplement Dates Manufacturer Received04/23/2019
05/21/2021
Supplement Dates FDA Received02/06/2020
05/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number3012977056-030619-001-R
Patient Sequence Number1
-
-