The anesthesia workstation was investigated on-site by our field service engineer.The nozzle units in the gas modules were replaced and this solved the reported flow transducer test failure.No further issues have been reported after this.The nozzle units were not returned for investigation.Evaluation of the received device logs confirm the reported flow transducer test failure during system checkout.The flow measured by the expiratory flow transducer was out of range during these tests.The error message was ¿expiratory flow transducer flow out of range¿.The error message indicates that the measured expiratory flow transducer flow, is outside the interval 90 % to 110 % of the total flow from all gas modules.Our conclusion, based on that no further issues have been reported, is that the nozzle units were the cause of the failure.The true root cause of the failure has not been determined as no parts were returned.
|