BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 20188
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/11/2016
Event Type  malfunction  
Event Description
It was reported that the stent dislodged.The 60-70% stenosed target lesion was located in a celiac artery.A 6.0mmx14mmx150cm express sd stent was advanced for treatment using a 7fr im guide catheter and a.018 guidewire.However, during procedure, it failed to cross the lesion.The physician attempted to pull the stent back into the guide catheter, but it would not go back.It was then attempted to remove the whole system but the stent came off of the balloon.A snare was used to remove the stent.The procedure was completed with a non-bsc device.No patient complications were reported and the patient was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8640258
MDR Text Key146063809
Report Number2134265-2019-05720
Device Sequence Number1
Product Code NIN
UDI-Device Identifier08714729484820
UDI-Public08714729484820
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0019179065
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/10/2019
Initial Date FDA Received05/24/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-