The product was not returned for evaluation; consequently, no product evaluation was performed.The complaint could not be confirmed.There are several different causes that could potentially be attributed to this issue.One potential reason is that the appearance of the intraocular lens (iol) may be changed by improper manipulation with forceps.Another factor may be that some iol marks may be caused by improper additive transfer (sourced by the loading chamber).Such marks are 100% biocompatible and typically disappear after several days or weeks.A definitive root cause cannot be determined.A device history record (dhr) review could not be completed as no lot number has been provided.Based upon the complaint trend review, the volume of complaints is within the acceptable control limits and no complaint trend is observed.The trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action is necessary at this time.
|