The introducer, catheter, stent covering and hemostasis tools were all returned for evaluation.The covering was completely detached from the stent but intact.The stent was implanted in the patient with no issues or complications.The hemostasis tools were undamaged.The bib catheter was passed through the d'vill introducer.A slight catch was felt at the transition between the flange and the sheath.The hemostasis valve was removed from the introducer.One side of the flange had not been heated enough, leaving part of the sheath tubing exposed.A 0.110" gauge pin was passed through the transition.The pin caught on the exposed tubing.The complaint was confirmed since the covering had detached from the stent.It appears that the covering caught on the exposed end of the sheath tubing at the bond between the flange and the tubing.It is likely that the covering would have remained attached if not for this issue.All covered cp stents are inspected in final qc to ensure each covering has (4) acceptable glue spots.A sample from each lot is test for covering strength.The sample from this lot had a covering strength of 2.41lbs, which is well above the 1.5lb acceptance criteria.
|