• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Break (1069)
Patient Problems Failure of Implant (1924); Inflammation (1932)
Event Date 04/30/2019
Event Type  Injury  
Manufacturer Narrative
The product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Results from the product history record review indicated the product met release criteria.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported that on the first postoperative day, the haptic was noticed to have been broken and it was sitting in front of the iris.The iol was exchanged for the same lens model and power on the first postoperative day.Additional information was provided indicating that in the surgeon¿s opinion the cartridge damage to iol amputation of haptic is what caused the event.There was no significant inflammation.
 
Manufacturer Narrative
Evaluation summary: the cartridge complaint product was not returned for analysis.The cartridge complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The qualified 20.5 diopter lens was returned.The reported haptic damage was observed.The lens had other damage consistent with an explant.The product investigation could not identify a root cause for the reported complaint.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key8660374
MDR Text Key146783421
Report Number1119421-2019-00801
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2019
Initial Date Manufacturer Received 05/07/2019
Initial Date FDA Received05/31/2019
Supplement Dates Manufacturer Received07/30/2019
Supplement Dates FDA Received08/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DISCOVISC; IOL DELIVERY SYSTEM; SN60WF 20.5 D
Patient Outcome(s) Required Intervention;
Patient Age72 YR
-
-