Model Number 9735669
Device Problems Computer Software Problem (1112); Data Problem (3196)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Concomitant medical product: sfw kit 9735736, stealth s8 ent.The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software tracking database.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation device being used outside of procedure, with no patient present.It was reported that while the site attempted to archive a patient, the received transfer failed.A manufacturer representative (rep) found that there was an exam tied to the patient with 0 slices.After deleting the extra exam, the issue was resolved and the site was able to export.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAIN CART 9735669 STEALTH S8 EM ENT
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8661614
MDR Text Key146813786
Report Number1723170-2019-02860
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169838918
UDI-Public00643169838918
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735669
Device Catalogue Number9735669
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received06/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-