• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number ROTAFLOW
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/17/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).Device not returned not eval.
 
Event Description
During patient treatment a "head error" occurred after each application of ultrasound couling agent, and then they rebooted after shutdown and the fault code disappeared.Complaint#: (b)(4).
 
Manufacturer Narrative
The service protocol has not been received despite several requests.This complaint will be closed without the service protocol.Therefore it cannot be assured in which technical status this device is currently.This complaint will be reopened and appropriate actions will then be taken if the service protocol is received and contains any further relevant information.And therefore no most probable root cause possible.For this error an nc was opened and investigated: in the course of the investigation of nc-17-06-011 all complaints were analyzed individually to look for patterns and causes.After evaluation of the complaints, the following defects are the most common: hot plug, sig error followed by head error, error message due to shaking / error message due to sensitivity, connection problems and hardware-error.The increase in complaints with the error "head error" is due to a user / application error.The actions have been addressed in the past to prevent application errors.Furthermore, the instructions for use of the rotaflow system see rotaflow system user manual, mcv-ga-10000703-de-11, contain detailed descriptions to prevent an "error head".In addition, the instruction manual describes how the user has to react in case of an error message so that the application can be continued if possible.Since there are several causes of errors that result in an error head (error head) message, no final root cause could be determined.It is noticeable that the evaluated error is largely due to a user error.Warning in the ifu regarding "hotplug" and label with the "hotplug" warning on the rfc have already been implemented in the past.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key8662338
MDR Text Key146808054
Report Number8010762-2019-00155
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup
Report Date 08/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW
Device Catalogue Number701046405
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/21/2019
Initial Date FDA Received06/03/2019
Supplement Dates Manufacturer Received05/21/2019
Supplement Dates FDA Received08/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-